Manufacturer of Zirconia Oxygen Analyzer - Hand-held Multi-Gas Analyzer – Med-link
Manufacturer of Zirconia Oxygen Analyzer - Hand-held Multi-Gas Analyzer – Med-link Detail:
Product features
1. This device is an anesthesia agent analyzer used to measure EtCO2, FiCO2, RR, EtN2O, FiN2O, EtAA, FiAA.
2. This monitor is suitable for all kinds of animals and may be applied in general ward, including, but not limited to ICU, CCU or ambulance and so forth.
Specifications
1. Main Unit’s Environment Requirement
Working: Temperature: 5℃~50℃;
Relative humindity: 0~95%;
Atmosphere Pressure: 70.0KPa~106.0KPa
Storage: Temperature: 0℃~70℃;
Relative humindity: 0~95%;
Atmosphere Pressure: 22.0KPa~120.0KPa
2. Power Specification
Input Voltage: 12V DC
Input Current: 2.0 A
3. Physical Specification
Main Unit
Weight: 0.65Kg
Dimension: 192mm x 106mm x 44mm
4. Hardware Specification
TFT Screen
Type: Colorful TFT LCD
Dimension: 5.0 inch
Battery
Quantity: 4
Model: Rechargeable lithium battery
Voltage: 3.7 V
Capacity: 2200mAh
Working time: 10 hours
Recharging time: 4 Hours
LED
Patient Alarm Indicator: Two colors: Yellow and Red
Sound Indicator
Loudspeaker: Play alarm voices
Interfaces
Power: 12VDC power socket x 1
USB: MINI USB socket x 1
5. Measurement Specification
Principle: NDIR single beam optics
Sampling Rate: 90mL/min, ±10mL/min
Initialization Time: Waveform displaying in 20 seconds
Range
CO2: 0~99 mmHg, 0~13 %
N2O: 0~100 VOL%
ISO: 0~6VOL%
ENF: 0~6VOL%
SEV: 0~8VOL%
RR: 2~150 bpm
Resolution
CO2: 0~40 mmHg ±2 mmHg
40 ~99 mmHg ±5% of reading
N2O: 0~100VOL%
±(2.0 vol% +5% of reading)
ISO: 0~6VOL%
±(0.3 vol% +2% of reading)
ENF: 0~6VOL%
±(0.3 vol% +2% of reading)
SEV: 0~8VOL%
±(0.3 vol% +2% of reading)
RR: ±1 bpm
Apnea Alarm Time: 20~60s
6. MAC value define
l 1.0MAC: under an atmospheric pressure condition, give a person or animal skin stimulation, 50% of people or animals not body dynamic response or escape reflection occurs, the inhalation anesthetics in alveolar concentration.
l To achieve 95% of people don't respond to stimuli, the MAC value should reach to 1.3.
l When MAC value is 0.4,most patients will wake up.
Anaesthetic agents
Enflurane: 1.68
Isoflurane: 1.16
Sevflurane: 1.71
Halothane: 0.75
N2O: 100%
Notice Desflurane’s MAC1.0 values differ with age
Age: 18-30 MAC1.0 7.25%
Age: 31-65 MAC1.0 6.0%
Product detail pictures:
Related Product Guide:
Rotation Medical Bioinductive Implant Promotes Consistent Rotator Cuff Healing, Mitigates Tear Progression | | Fetal Monitoring Probe
IRADIMED CORP (NASDAQ:IRMD) Big Money 2018 Q3 Sentiment at 1.95 | Temperature Sensor Probe
Our solutions are widely recognized and trusted by consumers and will meet up with constantly developing financial and social requires for Manufacturer of Zirconia Oxygen Analyzer - Hand-held Multi-Gas Analyzer – Med-link, The product will supply to all over the world, such as: Jeddah, Ethiopia, luzern, With excellent solutions, high quality service and sincere attitude of service, we ensure customer satisfaction and help customers create value for mutual benefit and create a win-win situation. Welcome customers all over the world to contact us or visit our company. We'll satisfy you with our qualified service!